Wine Analysis Home
pH Meter adjustments and calibration
Theory behind pH electodes and why they require calibrating
Temperature considerations
- pH meter adjustment
- It can be seen from the above equation that the sensing electrode's response to hydrogen ion activity (pH) is temperature dependent and needs to be allowed for.
- For this reason pH meters are either:
- fitted with a temperature adjustment control to manually set the pH meter to the temperature of solutions being measured
or
- fitted with a temperature probe (either as a separate entity or incorporated into the electrode arrangement) for automatic adjustments of temperatures involved.
- pH buffer and sample solutions' true pH
- When using standard pH buffers to calibrate your pH meter, the true pH of the buffer is temperature dependant and needs to be read of a chart or graph for that particular buffer's temperature.
NOTE: This is particularly relevant when pH buffers are stored in refrigerators.
- Similarly, sample pH measurements are also dependant on temperature, and should be recorded along side the temperature it was measured at.
NOTE: the pH meter temperature adjustment caters for the electrode's response to temperature, which is not to be confused with the temperature dependants of true pH values of solutions.
Some automated pH meters accomplish these measurements and calculations automatically. However, you need to ensure yourself that this is the case.
Electrode condition considerations (see electrode care)
- Reference Electrode
- The electrode potential of the reference electrode is dependant on the condition of the porous plug.
Contamination of this porous plug will alter the reference potential as well as slow down the electrode response.
- pH Electrode
- The electric potential of the special glass bulb of the pH electrode also varies through contamination, dehydration surface, abrasions and aging aspects of this glass bulb.
For the above reasons the pH electrode system needs to be calibrated.
1, 2 or 3 point calibrations
- One point calibration.
- As mentioned above, the pH measuring electrode responds in a linear fashion within the normal range of pH values expected close to that predicted by the Nernst equation.
- The slope of this curve (line) is calculated to equate to 59mv (millivolts) per pH unit change.
- To make sure that this is the case the pH meter may need to be adjusted via the pH 4.00, slope or sensitivity control.
- If the pH meter has recently had a 3, or 2 point calibration all that may be required is a one point calibration using a pH buffer close to the expected value of your sample solution to adjust the slope.
- A pH 3 or 4 buffer can be chosen for wine production pH measurements.
- A two (2) point calibration is normally recommended for each session
- Two point calibration.
Point one (1)
- At a true pH of 7.00 no electrical potential should exist across the pH electrode and read 0.00 millivolts .
- The pH meter may have to be adjusted so that this is the case.
- This adjustment should be the first calibration point as it will vary least when adjusting the slope with your second calibration point.
- A pH buffer close to pH 7.00 is used and the pH readout adjusted to read the true buffer pH.
- The adjustment can occur automatically or via the pH 7.00, zero, offset, gain, intercept or buffer adjustment control.
- The above adjustment control names or labels have their own logic depending on whether you are talking about the pH, voltage, graph, electronics or buffer aspects of the adjustments.
Point two (2)
- To set the slope, as discussed in the 1 point calibration above, use a pH buffer that ensure that the pH range expected is covered.
- A pH 4 buffer is usually chosen the buffer chosen to cover the expected pH throughout wine production.
- A 2 point calibration is normally recommended for each session
- Three point calibration.
- The third calibration of a three point calibration serves basically as a check that the pH meter has been zeroed correctly and the slope conforms to 57 mv per pH unit.
- A pH buffer close to the expected sample pH is usually chosen as the third point.
Go to the calibration procedure